Recall Z-2905-2026— TISSIUM SA

Class II · Ongoing

Class II FDA medical device recall by TISSIUM SA, initiated 2026-06-15. FDA's recall record names no specific 510(k) clearance for this event. Product: Brand Name: COAPTIUM CONNECT Product Name: COAPTIUM CONNECT Model/Catalog Number: NCOCO3010 Software Version: N/A Product Description: In situ polymerizing peripheral nerve repair device Component: No. Reason: Currently labeled shelf life of 36 months is not substantiated by real-time data. Devices will be relabeled to reflect revised shelf life of 26 months. The use of a medical device beyond its validated shelf life may result in perioperative breakage of the device rendering device unusable; worst-case outcome would be revision surgery..

Recalling firm
TISSIUM SA
Classification
Class II
Status
Ongoing
Initiated
2026-06-15
Firm location
Paris 12e Arrondissement, France

Product description

Brand Name: COAPTIUM CONNECT Product Name: COAPTIUM CONNECT Model/Catalog Number: NCOCO3010 Software Version: N/A Product Description: In situ polymerizing peripheral nerve repair device Component: No

Reason for recall

Currently labeled shelf life of 36 months is not substantiated by real-time data. Devices will be relabeled to reflect revised shelf life of 26 months. The use of a medical device beyond its validated shelf life may result in perioperative breakage of the device rendering device unusable; worst-case outcome would be revision surgery.

Data sourced from openFDA. This site is unofficial and independent of the FDA.