Recall Z-2892-2026— PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.

Class II · Ongoing

Class II FDA medical device recall by PHILIPS MEDICAL SYSTEMS NEDERLAND B.V., initiated 2026-06-17. It names 3 FDA 510(k) clearances: K200917, K181830, K172822. Product: Philips Azurion under the following System Descriptions and corresponding model numbers: 1. Azurion 3 M12; Model Numbers: 722063, 722221. 2. Azurion 3 M15; Model Numbers: 722064, 722222, 722280. 3. Azurion 5 M12; Model Number: 722227. 4. Azurion 5 M20; Model Numbers: 722228, 722281. 5. Azurion 7 B12; Model Numbers: 722067, 722225. 6. Azurion 7 B20; Model Numbers: 722068, 722226. 7. Azurion 7 M12; Model Numbers: 722078, 722223. 8. Azurion 7 M20; Model Numbers: 722079, 722224, 722282. Interventional Fluoroscopic X-Ray System.. Reason: Potential Loss of imaging functionality, loss of motorized movement, or loss of data due to unauthorized Hard Disk Drives..

Recalling firm
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Classification
Class II
Status
Ongoing
Initiated
2026-06-17
Firm location
Best, Netherlands

510(k) clearances named in this recall

3 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Philips Azurion under the following System Descriptions and corresponding model numbers: 1. Azurion 3 M12; Model Numbers: 722063, 722221. 2. Azurion 3 M15; Model Numbers: 722064, 722222, 722280. 3. Azurion 5 M12; Model Number: 722227. 4. Azurion 5 M20; Model Numbers: 722228, 722281. 5. Azurion 7 B12; Model Numbers: 722067, 722225. 6. Azurion 7 B20; Model Numbers: 722068, 722226. 7. Azurion 7 M12; Model Numbers: 722078, 722223. 8. Azurion 7 M20; Model Numbers: 722079, 722224, 722282. Interventional Fluoroscopic X-Ray System.

Reason for recall

Potential Loss of imaging functionality, loss of motorized movement, or loss of data due to unauthorized Hard Disk Drives.