Recall Z-2890-2020— Karl Storz Endoscopy

Class II · Ongoing

Class II FDA medical device recall by Karl Storz Endoscopy, initiated 2020-07-27. FDA's recall record names no specific 510(k) clearance for this event. Product: Flexible Intubation Fiberscope, Part: 11301AA1, with Instruction Manual: Z18441US - Product Usage: are intended to be used for examination and visualization of a patient s upper airway and for aiding placement of an endotracheal tube.. Reason: Updated sterilization instructions are being provided to ensure 1) a specific sterilization tray is not used, and 2) ethylene oxide sterilization is never used on certain reusable flexible intubation endoscopes with total lumen lengths greater than 850mm. Incompletely sterilized endoscopes might introduce infectious agents into patient airways, resulting in infection..

Recalling firm
Karl Storz Endoscopy
Classification
Class II
Status
Ongoing
Initiated
2020-07-27
Firm location
El Segundo, CA, United States

Product description

Flexible Intubation Fiberscope, Part: 11301AA1, with Instruction Manual: Z18441US - Product Usage: are intended to be used for examination and visualization of a patient s upper airway and for aiding placement of an endotracheal tube.

Reason for recall

Updated sterilization instructions are being provided to ensure 1) a specific sterilization tray is not used, and 2) ethylene oxide sterilization is never used on certain reusable flexible intubation endoscopes with total lumen lengths greater than 850mm. Incompletely sterilized endoscopes might introduce infectious agents into patient airways, resulting in infection.

Data sourced from openFDA. This site is unofficial and independent of the FDA.