Recall Z-2873-2020— Ortho-Clinical Diagnostics
Class II · Terminated
Class II FDA medical device recall by Ortho-Clinical Diagnostics, initiated 2020-07-13, terminated 2021-07-27. It names one FDA 510(k) clearance: K182063. Product: VITROS Chemistry Products TBIL Slides- 5 PACK/300 SLDS In vitro Diagnostic quantitatively measure total bilirubin (TBIL) concentration in serum and plasma, Product code: 8159931. Reason: Etrombopag Interference for VITROS Chemistry Products: TBIL Slides, VITROS BuBc Slides and VITROS XT TBIL-ALKP Slides.
- Recalling firm
- Ortho-Clinical Diagnostics
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2020-07-13
- Terminated
- 2021-07-27
- Firm location
- Rochester, NY, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
VITROS Chemistry Products TBIL Slides- 5 PACK/300 SLDS In vitro Diagnostic quantitatively measure total bilirubin (TBIL) concentration in serum and plasma, Product code: 8159931
Reason for recall
Etrombopag Interference for VITROS Chemistry Products: TBIL Slides, VITROS BuBc Slides and VITROS XT TBIL-ALKP Slides
Data sourced from openFDA. This site is unofficial and independent of the FDA.