Recall Z-2854-2026— Copan WASP

Class II · Ongoing

Class II FDA medical device recall by Copan WASP, initiated 2026-06-16. FDA's recall record names no specific 510(k) clearance for this event. Product: Brand Name: WASPLab Product Name: WASPLab Model/Catalog Number: W087 Software Version: WebApp version 2.33.1, 2.33.3 and 2.34.2 Product Description: WASPLab is an automated device intended to be used for the incubation and the digital imaging of agar culture plates streaked with microbiological specimens derived by the human body for microbiological investigations. WebApp software is the user interface of WASPLab device.. Reason: Due to a software issue within the WebApp that potentially prevents the user to overwrite the results..

Recalling firm
Copan WASP
Classification
Class II
Status
Ongoing
Initiated
2026-06-16
Firm location
Brescia, Italy

Product description

Brand Name: WASPLab Product Name: WASPLab Model/Catalog Number: W087 Software Version: WebApp version 2.33.1, 2.33.3 and 2.34.2 Product Description: WASPLab is an automated device intended to be used for the incubation and the digital imaging of agar culture plates streaked with microbiological specimens derived by the human body for microbiological investigations. WebApp software is the user interface of WASPLab device.

Reason for recall

Due to a software issue within the WebApp that potentially prevents the user to overwrite the results.