Recall Z-2846-2026— Pride Mobility Products Corp.
Class II · Ongoing
Class II FDA medical device recall by Pride Mobility Products Corp., initiated 2026-06-25. It names one FDA 510(k) clearance: K042612. Product: Brand Name: Quantum rehab Product Name: Q6 Edge 2.0 Model/Catalog Number: Q6 Edge 2.0 Software Version: 2.0.11, new version: 2.0.13 Product Description: Wheelchair, Powered (ITI), 890.3860 Component: Controller. Reason: If the "Remind Me Later" option is selected for user reminders, reminder pop-up messages may appear while the wheelchair is in use. If this occurs while driving the wheelchair, the wheelchair may perform an unexpected park brake stop, which may result in fall or injury. If this occurs while not driving the wheelchair, the control system may crash and remain in pre-operational mode..
- Recalling firm
- Pride Mobility Products Corp.
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2026-06-25
- Firm location
- Duryea, PA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Brand Name: Quantum rehab Product Name: Q6 Edge 2.0 Model/Catalog Number: Q6 Edge 2.0 Software Version: 2.0.11, new version: 2.0.13 Product Description: Wheelchair, Powered (ITI), 890.3860 Component: Controller
Reason for recall
If the "Remind Me Later" option is selected for user reminders, reminder pop-up messages may appear while the wheelchair is in use. If this occurs while driving the wheelchair, the wheelchair may perform an unexpected park brake stop, which may result in fall or injury. If this occurs while not driving the wheelchair, the control system may crash and remain in pre-operational mode.