Recall Z-2835-2018— Intuitive Surgical, Inc.
Class II · Ongoing
Class II FDA medical device recall by Intuitive Surgical, Inc., initiated 2018-05-17. It names one FDA 510(k) clearance: K131861. Product: The da Vinci Xi Surgical System (IS4000) Model Name: IS4000, Endoscope Controller. Model Number: 372601-16.. Reason: An Endoscope Controller has been improperly calibrated during manufacturing. This improper calibration may result in the Endoscope Controller showing surgical images that are dimmer, have color imbalance or have reduced fluorescence and grayscale when using Firefly. There is also potential for increased fogging of the endoscope..
- Recalling firm
- Intuitive Surgical, Inc.
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2018-05-17
- Firm location
- Sunnyvale, CA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
The da Vinci Xi Surgical System (IS4000) Model Name: IS4000, Endoscope Controller. Model Number: 372601-16.
Reason for recall
An Endoscope Controller has been improperly calibrated during manufacturing. This improper calibration may result in the Endoscope Controller showing surgical images that are dimmer, have color imbalance or have reduced fluorescence and grayscale when using Firefly. There is also potential for increased fogging of the endoscope.
Data sourced from openFDA. This site is unofficial and independent of the FDA.