Recall Z-2826-2018— Vyaire Medical

Class I · Terminated

Class I FDA medical device recall by Vyaire Medical, initiated 2018-06-20, terminated 2020-07-28. FDA's recall record names no specific 510(k) clearance for this event. Product: The Tri-Flo Subglottic Suction System. Reason: Vyaire Medical has discovered potential patient safety risk with associated with the Tri-Flow Subglottic Suction System. The distal soft tip of the catheter has been identified as having the potential to break off during patient use and result in aspiration of a solid foreign body..

Recalling firm
Vyaire Medical
Classification
Class I
Status
Terminated
Initiated
2018-06-20
Terminated
2020-07-28
Firm location
Mettawa, IL, United States

Product description

The Tri-Flo Subglottic Suction System

Reason for recall

Vyaire Medical has discovered potential patient safety risk with associated with the Tri-Flow Subglottic Suction System. The distal soft tip of the catheter has been identified as having the potential to break off during patient use and result in aspiration of a solid foreign body.