Recall Z-2824-2018— Sorin CRM SAS
Class II · Terminated
Class II FDA medical device recall by Sorin CRM SAS, initiated 2018-07-09, terminated 2024-09-06. FDA's recall record names no specific 510(k) clearance for this event. Product: Sorin Platinium CRT-D 1711 DF-1, Sterile EO. Implantable Cardiac Defibrillator and Resynchronization Therapy Defibrillator. Reason: Hardware configuration may lead to overconsumption, followed by loss of pacing and sensing capabilities. The device then cannot identify an arrhythmia that would require defibrillation shock therapy..
- Recalling firm
- Sorin CRM SAS
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2018-07-09
- Terminated
- 2024-09-06
- Firm location
- Clamart Cedex, France
Product description
Sorin Platinium CRT-D 1711 DF-1, Sterile EO. Implantable Cardiac Defibrillator and Resynchronization Therapy Defibrillator
Reason for recall
Hardware configuration may lead to overconsumption, followed by loss of pacing and sensing capabilities. The device then cannot identify an arrhythmia that would require defibrillation shock therapy.
Data sourced from openFDA. This site is unofficial and independent of the FDA.