Recall Z-2820-2015— Zimmer, Inc.

Class II · Terminated

Class II FDA medical device recall by Zimmer, Inc., initiated 2015-08-27, terminated 2016-03-15. FDA's recall record names no specific 510(k) clearance for this event. Product: Zimmer Guide Wires for orthopedic procedures.. Reason: Product labeling of these sterile guide wires (external carton label and patient record label) states a 10-year expiration date. Package testing supports a 5-year expiration date for these same guide wires..

Recalling firm
Zimmer, Inc.
Classification
Class II
Status
Terminated
Initiated
2015-08-27
Terminated
2016-03-15
Firm location
Warsaw, IN, United States

Product description

Zimmer Guide Wires for orthopedic procedures.

Reason for recall

Product labeling of these sterile guide wires (external carton label and patient record label) states a 10-year expiration date. Package testing supports a 5-year expiration date for these same guide wires.