Recall Z-2814-2015— Siemens Healthcare Diagnostics Inc

Class II · Terminated

Class II FDA medical device recall by Siemens Healthcare Diagnostics Inc, initiated 2015-08-24, terminated 2016-05-23. It names one FDA 510(k) clearance: K113216. Product: Siemens RAPIDPoint 500 Measurement Cartridge Lactate (400) tests Part Number: 10491448 Product Usage: The RAPIDPoint 500 system is designed for professional use in a point-of-care or laboratory environment. This system tests blood gases, electrolytes, metabolites, total hemoglobin, and hemoglobin derivatives in arterial, venous, and capillary whole blood samples. The following parameters are tested: pH, pC02, p02, Na+, K+, Ca++, Cl-, glucose, lactate, tHb, F02Hb, FCHb, FMetHb, FHHb, nBili. Reason: RAPIDPoint 500 measurement cartridge serial numbers 2517102517 through 2519000012, negative sodium bias on both QC and patient results.

Recalling firm
Siemens Healthcare Diagnostics Inc
Classification
Class II
Status
Terminated
Initiated
2015-08-24
Terminated
2016-05-23
Firm location
Norwood, MA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Siemens RAPIDPoint 500 Measurement Cartridge Lactate (400) tests Part Number: 10491448 Product Usage: The RAPIDPoint 500 system is designed for professional use in a point-of-care or laboratory environment. This system tests blood gases, electrolytes, metabolites, total hemoglobin, and hemoglobin derivatives in arterial, venous, and capillary whole blood samples. The following parameters are tested: pH, pC02, p02, Na+, K+, Ca++, Cl-, glucose, lactate, tHb, F02Hb, FCHb, FMetHb, FHHb, nBili

Reason for recall

RAPIDPoint 500 measurement cartridge serial numbers 2517102517 through 2519000012, negative sodium bias on both QC and patient results