Recall Z-2811-2026— Medtronic Perfusion Systems
Class II · Ongoing
Class II FDA medical device recall by Medtronic Perfusion Systems, initiated 2026-06-24. It names one FDA 510(k) clearance: K260195. Product: EOPA 3D Arterial Cannula 22FR, intended for use with cardiopulmonary bypass as an arterial return cannula, Model Number 78322. Reason: Pinhole leaks may occur in the wire-wound body during cardiopulmonary bypass. Potential risks associated with this condition may include hypovolemia, vessel laceration or perforation, hypotension, ischemia, and organ dysfunction..
- Recalling firm
- Medtronic Perfusion Systems
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2026-06-24
- Firm location
- Brooklyn Park, MN, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
EOPA 3D Arterial Cannula 22FR, intended for use with cardiopulmonary bypass as an arterial return cannula, Model Number 78322
Reason for recall
Pinhole leaks may occur in the wire-wound body during cardiopulmonary bypass. Potential risks associated with this condition may include hypovolemia, vessel laceration or perforation, hypotension, ischemia, and organ dysfunction.
Data sourced from openFDA. This site is unofficial and independent of the FDA.