Recall Z-2808-2026— Medtronic Perfusion Systems
Class II · Ongoing
Class II FDA medical device recall by Medtronic Perfusion Systems, initiated 2026-06-24. It names one FDA 510(k) clearance: K031037. Product: EOPA Elongated One-Piece Arterial Cannula DIL TIP 22FR, intended for use with cardiopulmonary bypass as an arterial return cannula, Model Number 77622. Reason: Pinhole leaks may occur in the wire-wound body during cardiopulmonary bypass. Potential risks associated with this condition may include hypovolemia, vessel laceration or perforation, hypotension, ischemia, and organ dysfunction..
- Recalling firm
- Medtronic Perfusion Systems
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2026-06-24
- Firm location
- Brooklyn Park, MN, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
EOPA Elongated One-Piece Arterial Cannula DIL TIP 22FR, intended for use with cardiopulmonary bypass as an arterial return cannula, Model Number 77622
Reason for recall
Pinhole leaks may occur in the wire-wound body during cardiopulmonary bypass. Potential risks associated with this condition may include hypovolemia, vessel laceration or perforation, hypotension, ischemia, and organ dysfunction.