Recall Z-2770-2017— Biomerieux Inc

Class II · Terminated

Class II FDA medical device recall by Biomerieux Inc, initiated 2017-05-16, terminated 2019-06-19. FDA's recall record names no specific 510(k) clearance for this event. Product: VITEK 2 Gram Negative test kits containing colistin (cs01n). Reason: The investigation demonstrated a high rate of very major errors (resistant isolates calling susceptible) with the VITEK¿ 2 AST-GN colistin (cs01n) compared to agar dilution (the reference method used for cs01n development) and compared to broth-microdilution (recommended method by EUCAST/CLSI)..

Recalling firm
Biomerieux Inc
Classification
Class II
Status
Terminated
Initiated
2017-05-16
Terminated
2019-06-19
Firm location
Hazelwood, MO, United States

Product description

VITEK 2 Gram Negative test kits containing colistin (cs01n)

Reason for recall

The investigation demonstrated a high rate of very major errors (resistant isolates calling susceptible) with the VITEK¿ 2 AST-GN colistin (cs01n) compared to agar dilution (the reference method used for cs01n development) and compared to broth-microdilution (recommended method by EUCAST/CLSI).