Recall Z-2768-2015— MedtestDx, Inc.

Class II · Terminated

Class II FDA medical device recall by MedtestDx, Inc., initiated 2015-05-15, terminated 2016-05-23. It names one FDA 510(k) clearance: K942458. Product: MedTest DX Control Reconstitution Fluid Kit configuration: 20 x 5 ml. Bottle: 5 ml amber glass Cap: 18 mm white seal plug seal rubber HC-DIL (Human Control Diluent) is reagent grade water filled into 5ml vials used to reconstitute Human controls HNC200 and HEC200. in vitro diagnostic.. Reason: Vial to vial variation in the fill volume that could affect control recovery issues and cause a control result to fall outside the suggested recovery ranges. An under-fill of the vials was found ranging as low as 4.63 ml. When used to reconstitute HNC/HEC, control values were shown to be out of range running on the high end..

Recalling firm
MedtestDx, Inc.
Classification
Class II
Status
Terminated
Initiated
2015-05-15
Terminated
2016-05-23
Firm location
Canton, MI, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

MedTest DX Control Reconstitution Fluid Kit configuration: 20 x 5 ml. Bottle: 5 ml amber glass Cap: 18 mm white seal plug seal rubber HC-DIL (Human Control Diluent) is reagent grade water filled into 5ml vials used to reconstitute Human controls HNC200 and HEC200. in vitro diagnostic.

Reason for recall

Vial to vial variation in the fill volume that could affect control recovery issues and cause a control result to fall outside the suggested recovery ranges. An under-fill of the vials was found ranging as low as 4.63 ml. When used to reconstitute HNC/HEC, control values were shown to be out of range running on the high end.