Recall Z-2760-2026— Ferno-Washington Inc
Class II · Ongoing
Class II FDA medical device recall by Ferno-Washington Inc, initiated 2026-06-19. FDA's recall record names no specific 510(k) clearance for this event. Product: Power F2 Powered Fastening System, Product No. 0567091 Powered fastening system intended for professional use only. Designed to load, unload, and secure a compatible wheeled stretcher inside a ground-based ambulance. Item no. 0567091. Reason: Product may experience accelerated roller wear and potential deformation due to interference between the roller assembly and rail. In the worst-case condition, rail mechanism may become lodged during operation and movement of the stretcher along the rail may no longer be possible through either powered or manual operation. May cause delays in transport of patient to healthcare facility..
- Recalling firm
- Ferno-Washington Inc
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2026-06-19
- Firm location
- Wilmington, OH, United States
Product description
Power F2 Powered Fastening System, Product No. 0567091 Powered fastening system intended for professional use only. Designed to load, unload, and secure a compatible wheeled stretcher inside a ground-based ambulance. Item no. 0567091
Reason for recall
Product may experience accelerated roller wear and potential deformation due to interference between the roller assembly and rail. In the worst-case condition, rail mechanism may become lodged during operation and movement of the stretcher along the rail may no longer be possible through either powered or manual operation. May cause delays in transport of patient to healthcare facility.
Data sourced from openFDA. This site is unofficial and independent of the FDA.