Recall Z-2758-2020— Siemens Medical Solutions USA, Inc
Class II · Terminated
Class II FDA medical device recall by Siemens Medical Solutions USA, Inc, initiated 2020-01-08, terminated 2022-06-21. It names one FDA 510(k) clearance: K133259. Product: SIEMENS Healthineers Ysio Max - Interventional Fluoroscopic X-Ray System, Model: 10762470 - Product Usage: is a device intended to visualize anatomical structures by converting an X-ray pattern into a visible image.. Reason: Two software issues (1) Using the override function in case of blocked system movements affecting Luminos dRF Max, Luminos Agile Max and Uroskop Omnia Max systems may cause collision of the system under operator control with obstacles or persons .(2) Incorrectly assigned image affecting Ysio Max, Luminos dRF Max and Luminos Agile Max systems with detector MAX Static causing incorrect base for diagnosis.
- Recalling firm
- Siemens Medical Solutions USA, Inc
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2020-01-08
- Terminated
- 2022-06-21
- Firm location
- Malvern, PA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
SIEMENS Healthineers Ysio Max - Interventional Fluoroscopic X-Ray System, Model: 10762470 - Product Usage: is a device intended to visualize anatomical structures by converting an X-ray pattern into a visible image.
Reason for recall
Two software issues (1) Using the override function in case of blocked system movements affecting Luminos dRF Max, Luminos Agile Max and Uroskop Omnia Max systems may cause collision of the system under operator control with obstacles or persons .(2) Incorrectly assigned image affecting Ysio Max, Luminos dRF Max and Luminos Agile Max systems with detector MAX Static causing incorrect base for diagnosis
Data sourced from openFDA. This site is unofficial and independent of the FDA.