Recall Z-2754-2017— Merge Healthcare, Inc.

Class II · Terminated

Class II FDA medical device recall by Merge Healthcare, Inc., initiated 2017-06-13, terminated 2020-09-17. It names one FDA 510(k) clearance: K110006. Product: Merge Eye Care PACS Viewer 5.2 Merge Eye Care PACS is a software solution for the display, management, archive, interface and integration of ophthalmic device reports, images and data. Merge Eye Care PACS is a software solution using databases for patient demographics, server and other systems of intercommunication with hospital systems via HL7 and DICOM to provide to clinicians a single image viewing and management solution of images imported from various contributing devices throughout the clinical environment. Merge Eye Care PACS provides the ability to review data from any network-connected computer, and is protected by appropriate security login which permits only authorized user access. Symphony uses 256 bit or greater encryption via secure socket layer (SSL) to assure a network environment which is secure.. Reason: When the user has not set up any user preference on the sorting order to render the study images (OS/OD/etc.) and reports, the ECP may get an incorrect study index. If the user attempts to delete an image, ECP may then read that image as OS when it is, in fact, OD..

Recalling firm
Merge Healthcare, Inc.
Classification
Class II
Status
Terminated
Initiated
2017-06-13
Terminated
2020-09-17
Firm location
Hartland, WI, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Merge Eye Care PACS Viewer 5.2 Merge Eye Care PACS is a software solution for the display, management, archive, interface and integration of ophthalmic device reports, images and data. Merge Eye Care PACS is a software solution using databases for patient demographics, server and other systems of intercommunication with hospital systems via HL7 and DICOM to provide to clinicians a single image viewing and management solution of images imported from various contributing devices throughout the clinical environment. Merge Eye Care PACS provides the ability to review data from any network-connected computer, and is protected by appropriate security login which permits only authorized user access. Symphony uses 256 bit or greater encryption via secure socket layer (SSL) to assure a network environment which is secure.

Reason for recall

When the user has not set up any user preference on the sorting order to render the study images (OS/OD/etc.) and reports, the ECP may get an incorrect study index. If the user attempts to delete an image, ECP may then read that image as OS when it is, in fact, OD.