Recall Z-2747-2020— Siemens Healthcare Diagnostics, Inc.
Class II · Terminated
Class II FDA medical device recall by Siemens Healthcare Diagnostics, Inc., initiated 2020-06-30, terminated 2020-11-17. It names one FDA 510(k) clearance: K071466. Product: DCA Vantage Handheld Barcode Scanner - Zebra Model - Model # DS4308 - HC0015BZZWW external accessory to DCA Vantage Analyzer US. Reason: If DCA Vantage Analyzer is configured to run Code 39 with check digit, the scanner is not confirming the integrity of read barcode data using the check digit. Also, if DCA Vantage Analyzer is configured to run Code 39 with check digit and an operator erroneously scans a Code 39 barcode without a check digit, the scanner incorrectly sends the code and the last character is removed..
- Recalling firm
- Siemens Healthcare Diagnostics, Inc.
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2020-06-30
- Terminated
- 2020-11-17
- Firm location
- Newark, DE, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
DCA Vantage Handheld Barcode Scanner - Zebra Model - Model # DS4308 - HC0015BZZWW external accessory to DCA Vantage Analyzer US
Reason for recall
If DCA Vantage Analyzer is configured to run Code 39 with check digit, the scanner is not confirming the integrity of read barcode data using the check digit. Also, if DCA Vantage Analyzer is configured to run Code 39 with check digit and an operator erroneously scans a Code 39 barcode without a check digit, the scanner incorrectly sends the code and the last character is removed.