Recall Z-2744-2018— ConvaTec, Inc

Class II · Terminated

Class II FDA medical device recall by ConvaTec, Inc, initiated 2018-07-03, terminated 2023-01-24. FDA's recall record names no specific 510(k) clearance for this event. Product: AbViser AutoValve IAP (intra- abdominal pressure) Monitoring Device, Patient mount: a) REF ABV300, b) REF ABV301 AbViser is a sterile non-invasive disposable intra- abdominal pressure monitoring device containing aspiration tubing, infusion tubing, valves and optional pressure transducers for the measurement of intra- abdominal bladder pressure. The device attaches directly to the patient s existing urinary catheter/drain system providing both an enclosed fluid path for infusing fluid into the bladder catheter in order to monitor the hydrostatic pressure in the bladder.. Reason: Bubble Leak Testing on AbViser AutoValve IAP Monitoring Devices without Transducer confirmed the potential for a pinhole breach in the sterile barrier..

Recalling firm
ConvaTec, Inc
Classification
Class II
Status
Terminated
Initiated
2018-07-03
Terminated
2023-01-24
Firm location
Greensboro, NC, United States

Product description

AbViser AutoValve IAP (intra- abdominal pressure) Monitoring Device, Patient mount: a) REF ABV300, b) REF ABV301 AbViser is a sterile non-invasive disposable intra- abdominal pressure monitoring device containing aspiration tubing, infusion tubing, valves and optional pressure transducers for the measurement of intra- abdominal bladder pressure. The device attaches directly to the patient s existing urinary catheter/drain system providing both an enclosed fluid path for infusing fluid into the bladder catheter in order to monitor the hydrostatic pressure in the bladder.

Reason for recall

Bubble Leak Testing on AbViser AutoValve IAP Monitoring Devices without Transducer confirmed the potential for a pinhole breach in the sterile barrier.

Data sourced from openFDA. This site is unofficial and independent of the FDA.