Recall Z-2742-2020— OrthoPediatrics Corp

Class II · Ongoing

Class II FDA medical device recall by OrthoPediatrics Corp, initiated 2020-07-06. It names one FDA 510(k) clearance: K150600. Product: Orthopedics 6.0MM ROD, COCR, SINGLE HEX, 500MM LENGTH-The RESPONSE" 5.5/6.0 Spine System is intended for immobilization and stabilization of the posterior, non-cervical spine in skeletally mature patients as an adjunct to fusion for the following indications: degenerative disc disease Model Number: 00-1300-6053. Reason: Product may have incorrect laser etched alignment lines, the non-conforming laser etch may have compromised strength, lead to unintended fracture of the device.

Recalling firm
OrthoPediatrics Corp
Classification
Class II
Status
Ongoing
Initiated
2020-07-06
Firm location
Warsaw, IN, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Orthopedics 6.0MM ROD, COCR, SINGLE HEX, 500MM LENGTH-The RESPONSE" 5.5/6.0 Spine System is intended for immobilization and stabilization of the posterior, non-cervical spine in skeletally mature patients as an adjunct to fusion for the following indications: degenerative disc disease Model Number: 00-1300-6053

Reason for recall

Product may have incorrect laser etched alignment lines, the non-conforming laser etch may have compromised strength, lead to unintended fracture of the device

Data sourced from openFDA. This site is unofficial and independent of the FDA.