Recall Z-2741-2026— Maquet Cardiovascular, LLC

Class II · Ongoing

Class II FDA medical device recall by Maquet Cardiovascular, LLC, initiated 2026-06-02. Product: The Vasoview Hemopro 2 Endoscopic Vessel Harvesting System (Model VH-4000; UDI 00607567700406) is a sterile, single-use Class II device that was cleared for marketing by FDA under 510(k) number K101274.. Reason: Getinge identified that Vasoview Hemopro 2 devices from Lot 3000547797 were exposed to an out-of-specification sterilization dose during a gamma sterilization run (Run ID 1183-28782A) performed by STERIS at its Whippany, NJ facility between March 31 and April 1 of 2026..

Recalling firm
Maquet Cardiovascular, LLC
Classification
Class II
Status
Ongoing
Initiated
2026-06-02
Firm location
Wayne, NJ, United States

Product description

The Vasoview Hemopro 2 Endoscopic Vessel Harvesting System (Model VH-4000; UDI 00607567700406) is a sterile, single-use Class II device that was cleared for marketing by FDA under 510(k) number K101274.

Reason for recall

Getinge identified that Vasoview Hemopro 2 devices from Lot 3000547797 were exposed to an out-of-specification sterilization dose during a gamma sterilization run (Run ID 1183-28782A) performed by STERIS at its Whippany, NJ facility between March 31 and April 1 of 2026.