Recall Z-2737-2016— Philips Medical Systems (Cleveland) Inc
Class II · Terminated
Class II FDA medical device recall by Philips Medical Systems (Cleveland) Inc, initiated 2016-05-25, terminated 2018-07-02. It names one FDA 510(k) clearance: K080927. Product: BrightView XCT Upgrade Model 882454. Reason: The firm became aware of an issue where during a non-Auto Body Contouring (ABC), non-circular, rel-180 scan, the detector may come in contact with the patient. The issue was caused by a software defect in the workflow setup script..
- Recalling firm
- Philips Medical Systems (Cleveland) Inc
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2016-05-25
- Terminated
- 2018-07-02
- Firm location
- Cleveland, OH, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
BrightView XCT Upgrade Model 882454
Reason for recall
The firm became aware of an issue where during a non-Auto Body Contouring (ABC), non-circular, rel-180 scan, the detector may come in contact with the patient. The issue was caused by a software defect in the workflow setup script.