Recall Z-2733-2018— INO therapeutics LLC/dba ikaria

Class II · Terminated

Class II FDA medical device recall by INO therapeutics LLC/dba ikaria, initiated 2018-07-09, terminated 2020-11-30. It names one FDA 510(k) clearance: K131686. Product: Mallinckrodt Pharmaceuticals INOflo DS, NO Gas Control System. Product Usage; Device Classification: The INOmeter is an accessory to the nitric oxide delivery device which is used for the delivery of inhaled nitric oxide to patient. An INOmeter is affixed to each cylinder of drug product. Functionality of the INOmeter: The INOmeter is affixed on top of each INOflo cylinder valve in Japan. When the healthcare provider opens the INOmeter the compressed gas cylinder valve opens to begin flow of the nitric oxide in nitrogen compressed gas. The nitric oxide gas is administered to a patient via a nitric oxide delivery device. The amount of time that gas flow administered to a patient is captured in the INOmeter. Each INOmeter is programmed to include the drug product expiration date as listed on the INOflo drug product label. The INOmeter electronically communicates to the nitric oxide delivery device via infrared technology by way of the iButton and in so communicates the expiration date of the corresponding cylinder of INOflo drug product to the delivery device. The delivery device evaluates and compares the expiration date to the current date of therapy administration. If the drug product expiration date is prior to the current date, an error and alarm is registered by the nitric oxide delivery device and administration of the expired drug to the patient is stopped after 120 seconds. If this occurs, the healthcare provider can operate the nitric oxide delivery device manually and deliver the INOflo as indicated in the Instructions For Use (IFU). The INOflo delivery device has the functionality to administer INOflo in manual mode.. Reason: INOmeters affixed to the INOflo cylinders in Lot 17FP-0570 were programmed with an incorrect expiration date (11/2017). For Lot 17FP-0570, the label affixed to each INOflo cylinder has the correct expiration date noted (11/2020)..

Recalling firm
INO therapeutics LLC/dba ikaria
Classification
Class II
Status
Terminated
Initiated
2018-07-09
Terminated
2020-11-30
Firm location
Port Allen, LA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Mallinckrodt Pharmaceuticals INOflo DS, NO Gas Control System. Product Usage; Device Classification: The INOmeter is an accessory to the nitric oxide delivery device which is used for the delivery of inhaled nitric oxide to patient. An INOmeter is affixed to each cylinder of drug product. Functionality of the INOmeter: The INOmeter is affixed on top of each INOflo cylinder valve in Japan. When the healthcare provider opens the INOmeter the compressed gas cylinder valve opens to begin flow of the nitric oxide in nitrogen compressed gas. The nitric oxide gas is administered to a patient via a nitric oxide delivery device. The amount of time that gas flow administered to a patient is captured in the INOmeter. Each INOmeter is programmed to include the drug product expiration date as listed on the INOflo drug product label. The INOmeter electronically communicates to the nitric oxide delivery device via infrared technology by way of the iButton and in so communicates the expiration date of the corresponding cylinder of INOflo drug product to the delivery device. The delivery device evaluates and compares the expiration date to the current date of therapy administration. If the drug product expiration date is prior to the current date, an error and alarm is registered by the nitric oxide delivery device and administration of the expired drug to the patient is stopped after 120 seconds. If this occurs, the healthcare provider can operate the nitric oxide delivery device manually and deliver the INOflo as indicated in the Instructions For Use (IFU). The INOflo delivery device has the functionality to administer INOflo in manual mode.

Reason for recall

INOmeters affixed to the INOflo cylinders in Lot 17FP-0570 were programmed with an incorrect expiration date (11/2017). For Lot 17FP-0570, the label affixed to each INOflo cylinder has the correct expiration date noted (11/2020).