Recall Z-2729-2024— Stryker Corporation

Class II · Completed

Class II FDA medical device recall by Stryker Corporation, initiated 2024-07-24. It names one FDA 510(k) clearance: K210377. Product: Stryker iBur 5.0mm Coarse Diamond Round, Distal Bend intended to cut bone in the following manner: drilling, reaming, decorticating, shaping, dissecting, shaving, and smoothing for the following medical applications: Neuro; Spine; Ear, Nose, and Throat (ENT)/ Otorhinolaryngology; and Endoscopic applications Ref: 8431-013-050DC. Reason: Potential for the product to exhibit temperatures higher than specification where the bur shank meets the distal bushing. Temperatures higher than specification may lead to the potential for minor tissue/structure damage or tissue/structure damage from thermal injury which may need medical/surgical intervention..

Recalling firm
Stryker Corporation
Classification
Class II
Status
Completed
Initiated
2024-07-24
Firm location
Portage, MI, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Stryker iBur 5.0mm Coarse Diamond Round, Distal Bend intended to cut bone in the following manner: drilling, reaming, decorticating, shaping, dissecting, shaving, and smoothing for the following medical applications: Neuro; Spine; Ear, Nose, and Throat (ENT)/ Otorhinolaryngology; and Endoscopic applications Ref: 8431-013-050DC

Reason for recall

Potential for the product to exhibit temperatures higher than specification where the bur shank meets the distal bushing. Temperatures higher than specification may lead to the potential for minor tissue/structure damage or tissue/structure damage from thermal injury which may need medical/surgical intervention.

Data sourced from openFDA. This site is unofficial and independent of the FDA.