Recall Z-2727-2026— GE Medical Systems, LLC

Class II · Ongoing

Class II FDA medical device recall by GE Medical Systems, LLC, initiated 2026-06-05. Product: Optima XR220amx, Mobile Digital-Ready Radiographic System. Reason: GE HealthCare has become aware that in certain Optima XR200/220/240amx systems that have had a column replacement completed, functional testing of the safety pawl may not have been performed to confirm proper assembly. If a failure of the internal counterpoise cable occurs, the Column Safety mechanism will not work, causing the arm with X-ray tube and collimator to drop downward on the column track. This could result in patient injury during an exam..

Recalling firm
GE Medical Systems, LLC
Classification
Class II
Status
Ongoing
Initiated
2026-06-05
Firm location
Waukesha, WI, United States

Product description

Optima XR220amx, Mobile Digital-Ready Radiographic System

Reason for recall

GE HealthCare has become aware that in certain Optima XR200/220/240amx systems that have had a column replacement completed, functional testing of the safety pawl may not have been performed to confirm proper assembly. If a failure of the internal counterpoise cable occurs, the Column Safety mechanism will not work, causing the arm with X-ray tube and collimator to drop downward on the column track. This could result in patient injury during an exam.