Recall Z-2721-2016— Elekta, Inc.

Class II · Terminated

Class II FDA medical device recall by Elekta, Inc., initiated 2016-07-08, terminated 2017-11-24. It names one FDA 510(k) clearance: K051932. Product: Medical Linear Accelerator. Reason: There have been instances in hospitals where the joint between the drive gear and the gantry base drive wheel has become loose, leading to fatigue failure of some of the securing bolts..

Recalling firm
Elekta, Inc.
Classification
Class II
Status
Terminated
Initiated
2016-07-08
Terminated
2017-11-24
Firm location
Atlanta, GA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Medical Linear Accelerator

Reason for recall

There have been instances in hospitals where the joint between the drive gear and the gantry base drive wheel has become loose, leading to fatigue failure of some of the securing bolts.