Recall Z-2718-2024— Treace Medical Concepts, Inc.

Class II · Ongoing

Class II FDA medical device recall by Treace Medical Concepts, Inc., initiated 2024-06-02. It names one FDA 510(k) clearance: K240173. Product: Nanoplasty 3D Bunion Correction System Implant, Model Number: SK57. System contains the 01-00066 1st Metatarsal Plate as a component.. Reason: Potential issue causing a higher frequency of interference with locking screws and/or inserter rods than anticipated, resulting in screws stripping and/or not fully seating in the plates and/or interface issues between inserter rods and plates..

Recalling firm
Treace Medical Concepts, Inc.
Classification
Class II
Status
Ongoing
Initiated
2024-06-02
Firm location
Ponte Vedra, FL, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Nanoplasty 3D Bunion Correction System Implant, Model Number: SK57. System contains the 01-00066 1st Metatarsal Plate as a component.

Reason for recall

Potential issue causing a higher frequency of interference with locking screws and/or inserter rods than anticipated, resulting in screws stripping and/or not fully seating in the plates and/or interface issues between inserter rods and plates.