Recall Z-2708-2014— Dako North America Inc.

Class II · Terminated

Class II FDA medical device recall by Dako North America Inc., initiated 2014-08-29, terminated 2015-03-03. FDA's recall record names no specific 510(k) clearance for this event. Product: Dako Autostainer Link 48 with software version Dako Link 4.0.3, Model No. AS480 Product Usage: For in vitro diagnostic use. The Dako Autostainer Link 48 is an automated slide processing system compatible with reagents for the staining of paraffin-embedded and frozen tissue sections, cytospins, cell smears and fine needle aspirates. The system is designed to automate manual staining methods routinely used in immunohistochemistry and cytochemistry, enabling the transfer of established protocols from the bench to the Autostainer Link 48 automated workstations operating either singly or in networked configurations.. Reason: Dako is recalling the Dako Link 4.0.3 software on the Dako Autostainer Link 48 because Dako has determined that an upgrade to this version of the software may cause premature rejection of reagent bottles or inadequate aspiration of reagent if the instrument has also had a preventative maintenance or waste pump replacement, and the baseplate is not completely aligned..

Recalling firm
Dako North America Inc.
Classification
Class II
Status
Terminated
Initiated
2014-08-29
Terminated
2015-03-03
Firm location
Carpinteria, CA, United States

Product description

Dako Autostainer Link 48 with software version Dako Link 4.0.3, Model No. AS480 Product Usage: For in vitro diagnostic use. The Dako Autostainer Link 48 is an automated slide processing system compatible with reagents for the staining of paraffin-embedded and frozen tissue sections, cytospins, cell smears and fine needle aspirates. The system is designed to automate manual staining methods routinely used in immunohistochemistry and cytochemistry, enabling the transfer of established protocols from the bench to the Autostainer Link 48 automated workstations operating either singly or in networked configurations.

Reason for recall

Dako is recalling the Dako Link 4.0.3 software on the Dako Autostainer Link 48 because Dako has determined that an upgrade to this version of the software may cause premature rejection of reagent bottles or inadequate aspiration of reagent if the instrument has also had a preventative maintenance or waste pump replacement, and the baseplate is not completely aligned.

Data sourced from openFDA. This site is unofficial and independent of the FDA.