Recall Z-2707-2017— Merge Healthcare, Inc.
Class II · Terminated
Class II FDA medical device recall by Merge Healthcare, Inc., initiated 2016-03-10, terminated 2020-03-06. It names one FDA 510(k) clearance: K082421. Product: The Schiller PB 1000, also known as Merge Hemo Patient Data Module (PDM), a major component of the Merge Hemo software.. Reason: If the PDM is utilized to output a pressure for computing fractional flow reserve (FFR), the analog output may not reflect the true representation of the selected pressure waveform being provided by the PDM to the Link Assembly which provides the analog output..
- Recalling firm
- Merge Healthcare, Inc.
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2016-03-10
- Terminated
- 2020-03-06
- Firm location
- Hartland, WI, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
The Schiller PB 1000, also known as Merge Hemo Patient Data Module (PDM), a major component of the Merge Hemo software.
Reason for recall
If the PDM is utilized to output a pressure for computing fractional flow reserve (FFR), the analog output may not reflect the true representation of the selected pressure waveform being provided by the PDM to the Link Assembly which provides the analog output.
Data sourced from openFDA. This site is unofficial and independent of the FDA.