Recall Z-2707-2017— Merge Healthcare, Inc.

Class II · Terminated

Class II FDA medical device recall by Merge Healthcare, Inc., initiated 2016-03-10, terminated 2020-03-06. It names one FDA 510(k) clearance: K082421. Product: The Schiller PB 1000, also known as Merge Hemo Patient Data Module (PDM), a major component of the Merge Hemo software.. Reason: If the PDM is utilized to output a pressure for computing fractional flow reserve (FFR), the analog output may not reflect the true representation of the selected pressure waveform being provided by the PDM to the Link Assembly which provides the analog output..

Recalling firm
Merge Healthcare, Inc.
Classification
Class II
Status
Terminated
Initiated
2016-03-10
Terminated
2020-03-06
Firm location
Hartland, WI, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

The Schiller PB 1000, also known as Merge Hemo Patient Data Module (PDM), a major component of the Merge Hemo software.

Reason for recall

If the PDM is utilized to output a pressure for computing fractional flow reserve (FFR), the analog output may not reflect the true representation of the selected pressure waveform being provided by the PDM to the Link Assembly which provides the analog output.