Recall Z-2697-2020— Philips North America, LLC

Class II · Terminated

Class II FDA medical device recall by Philips North America, LLC, initiated 2020-06-30, terminated 2024-01-25. It names one FDA 510(k) clearance: K150310. Product: IntelliVue MX700 patient monitor, model no. 865241. Reason: Some Intellivue MX700/MX800 Patient Monitors were delivered with defective power supplies. The defective power supplies might energize the metal parts of monitor if the grounding instructions were not followed, and may also cause the monitor to stop operating or prevent it from being switched on..

Recalling firm
Philips North America, LLC
Classification
Class II
Status
Terminated
Initiated
2020-06-30
Terminated
2024-01-25
Firm location
Andover, MA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

IntelliVue MX700 patient monitor, model no. 865241

Reason for recall

Some Intellivue MX700/MX800 Patient Monitors were delivered with defective power supplies. The defective power supplies might energize the metal parts of monitor if the grounding instructions were not followed, and may also cause the monitor to stop operating or prevent it from being switched on.