Recall Z-2693-2017— St. Jude Medical, Inc.
Class II · Terminated
Class II FDA medical device recall by St. Jude Medical, Inc., initiated 2017-06-02, terminated 2022-08-24. FDA's recall record names no specific 510(k) clearance for this event. Product: Infinity 5, Implantable Pulse Generator, REF 6660, Rx only, STERILE EO, Product Usage: The Infinity IPG is an electronic device designed to be connected to one or two extensions. It is powered by a hermetically sealed battery within a titanium case and uses microelectronic circuitry to generate constant-current electrical stimulation. The IPG is conductive on all sides, which allows the IPG case (also called a can) to be used as an anode for monopolar stimulation. The IPG communicates wirelessly with system programmers and controllers, and IPGs are available in small and large sizes to accommodate different power needs.. Reason: The firm discovered a risk of loss of Neuromodulation therapy due to exposure of the IPG to monopolar electrosurgical devices (ESU) during medical procedures..
- Recalling firm
- St. Jude Medical, Inc.
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2017-06-02
- Terminated
- 2022-08-24
- Firm location
- Plano, TX, United States
Product description
Infinity 5, Implantable Pulse Generator, REF 6660, Rx only, STERILE EO, Product Usage: The Infinity IPG is an electronic device designed to be connected to one or two extensions. It is powered by a hermetically sealed battery within a titanium case and uses microelectronic circuitry to generate constant-current electrical stimulation. The IPG is conductive on all sides, which allows the IPG case (also called a can) to be used as an anode for monopolar stimulation. The IPG communicates wirelessly with system programmers and controllers, and IPGs are available in small and large sizes to accommodate different power needs.
Reason for recall
The firm discovered a risk of loss of Neuromodulation therapy due to exposure of the IPG to monopolar electrosurgical devices (ESU) during medical procedures.
Data sourced from openFDA. This site is unofficial and independent of the FDA.