Recall Z-2692-2024— Abbott Diabetes Care, Inc.

Class I · Ongoing

Class I FDA medical device recall by Abbott Diabetes Care, Inc., initiated 2024-07-24. It names one FDA 510(k) clearance: K212132. Product: FreeStyle Libre 3 Sensors, REF: 72081-01, component of the FreeStyle Libre 3 Continuous Glucose Monitoring System. Reason: Users of the FreeStyle Libre 3 sensors reported situations where they were receiving erroneously high glucose results. The inaccurate higher glucose values may lead to users calculating higher insulin bolus correctional doses based on inaccurate CGM glucose data. The inaccurate higher glucose values may also contribute to missed or delayed recognition of hypoglycemia, when the sensor incorrectly reports normal or even high glucose values when the actual glucose is trending low or is low. If the hypoglycemia missed by the device occurs acutely, a user may not be able to confirm the actual glucose with a self-monitored blood glucose (BG) and/or administer prompt intervention to raise glucose levels. The over-delivery of insulin and/or the missed or delayed detection of impending or overt hypoglycemia in the worst-case scenario may lead to severe hypoglycemia/hypoglycemic crisis with significant adverse health consequences. Such consequences include central nervous system dysfunction, loss of consciousness, seizure activity and may lead to coma, permanent neurological damage, and death..

Recalling firm
Abbott Diabetes Care, Inc.
Classification
Class I
Status
Ongoing
Initiated
2024-07-24
Firm location
Alameda, CA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

FreeStyle Libre 3 Sensors, REF: 72081-01, component of the FreeStyle Libre 3 Continuous Glucose Monitoring System

Reason for recall

Users of the FreeStyle Libre 3 sensors reported situations where they were receiving erroneously high glucose results. The inaccurate higher glucose values may lead to users calculating higher insulin bolus correctional doses based on inaccurate CGM glucose data. The inaccurate higher glucose values may also contribute to missed or delayed recognition of hypoglycemia, when the sensor incorrectly reports normal or even high glucose values when the actual glucose is trending low or is low. If the hypoglycemia missed by the device occurs acutely, a user may not be able to confirm the actual glucose with a self-monitored blood glucose (BG) and/or administer prompt intervention to raise glucose levels. The over-delivery of insulin and/or the missed or delayed detection of impending or overt hypoglycemia in the worst-case scenario may lead to severe hypoglycemia/hypoglycemic crisis with significant adverse health consequences. Such consequences include central nervous system dysfunction, loss of consciousness, seizure activity and may lead to coma, permanent neurological damage, and death.