Recall Z-2680-2019— Medtronic Vascular

Class I · Completed

Class I FDA medical device recall by Medtronic Vascular, initiated 2019-03-15, terminated 2024-04-26. It names 2 FDA 510(k) clearances: K132673, K042489. Product: SHERPA NX ACTIVE GUIDING CATHETER, 6F HSII, .070", REF SA6HSII. for cardiovascular use. Reason: There is a potential for extensive loss of primary segment material exposing underlying stainless-steel braid wires on a subset of 6F Sherpa catheters..

Recalling firm
Medtronic Vascular
Classification
Class I
Status
Completed
Initiated
2019-03-15
Terminated
2024-04-26
Firm location
Danvers, MA, United States

510(k) clearances named in this recall

2 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

SHERPA NX ACTIVE GUIDING CATHETER, 6F HSII, .070", REF SA6HSII. for cardiovascular use

Reason for recall

There is a potential for extensive loss of primary segment material exposing underlying stainless-steel braid wires on a subset of 6F Sherpa catheters.