Recall Z-2673-2016— Nakanishi Inc.

Class II · Terminated

Class II FDA medical device recall by Nakanishi Inc., initiated 2016-07-08, terminated 2018-06-25. It names one FDA 510(k) clearance: K073678. Product: Diamond Coated Tips and Burs are accessories to the air powered Bone drill (VARIOSURG with and without light) The intended use of the tips is based on the parent device. For the VarioSurg, in addition to the scaling functions, diamond coated tips are used for socket lift, sinus lift, and implant preparation.. Reason: Through an error, the labeling failed to include the symbol or other text indicating that the diamond coated tips were single use only. A contributing factor to this error is that in other markets such as Japan, the diamond coated tips are multiple use and are not limited to single use only. The tips are stainless steel with a coating of fine diamond powder. The tips are sold non-sterile and are to be sterilized prior to first use. The tips were cleared as single use only. This problem was discovered during an ongoing, proactive, quality system improvement plan that was initiated by Nakanishi in October 2014. Because the correct instructions for sterilization are not available, it is possible that the devices might not be sterile prior to use. The absence of the single use statement introduces a risk of cross-contamination if the tips are reprocessed and reused..

Recalling firm
Nakanishi Inc.
Classification
Class II
Status
Terminated
Initiated
2016-07-08
Terminated
2018-06-25
Firm location
Kanuma, N/A, Japan

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Diamond Coated Tips and Burs are accessories to the air powered Bone drill (VARIOSURG with and without light) The intended use of the tips is based on the parent device. For the VarioSurg, in addition to the scaling functions, diamond coated tips are used for socket lift, sinus lift, and implant preparation.

Reason for recall

Through an error, the labeling failed to include the symbol or other text indicating that the diamond coated tips were single use only. A contributing factor to this error is that in other markets such as Japan, the diamond coated tips are multiple use and are not limited to single use only. The tips are stainless steel with a coating of fine diamond powder. The tips are sold non-sterile and are to be sterilized prior to first use. The tips were cleared as single use only. This problem was discovered during an ongoing, proactive, quality system improvement plan that was initiated by Nakanishi in October 2014. Because the correct instructions for sterilization are not available, it is possible that the devices might not be sterile prior to use. The absence of the single use statement introduces a risk of cross-contamination if the tips are reprocessed and reused.

Data sourced from openFDA. This site is unofficial and independent of the FDA.