Recall Z-2671-2014— ORTHO-CLINICAL DIAGNOSTICS

Class II · Terminated

Class II FDA medical device recall by ORTHO-CLINICAL DIAGNOSTICS, initiated 2014-08-06, terminated 2017-04-14. FDA's recall record names no specific 510(k) clearance for this event. Product: VITROS Immunodiagnostic Products Anti-HCV Reagent Pack for in vitro qualitative detection of immunoglobulin G antibody to hepatitis C virus (anti-HCV) in human serum and plasma. REF 680 2454, IVD -- Ortho-Clinical Diagnostics High Wycombe UK. Note that one sales unit has 5 Reagent Packs contained within it. Each Reagent Pack can test for 100 samples (patient, Quality Control, etc.).. Reason: VITROS Immunodiagnostic Products Anti-HCV kit lot 9090 has been observed producing sporadic lower than expected VITROS Anti-HCV test results due to an approximate signal loss of up to 66.2% which can result in unexpected negative test results being obtained for reactive sample fluids (quality control and patient samples)..

Recalling firm
ORTHO-CLINICAL DIAGNOSTICS
Classification
Class II
Status
Terminated
Initiated
2014-08-06
Terminated
2017-04-14
Firm location
Bridgend, N/A, United Kingdom

Product description

VITROS Immunodiagnostic Products Anti-HCV Reagent Pack for in vitro qualitative detection of immunoglobulin G antibody to hepatitis C virus (anti-HCV) in human serum and plasma. REF 680 2454, IVD -- Ortho-Clinical Diagnostics High Wycombe UK. Note that one sales unit has 5 Reagent Packs contained within it. Each Reagent Pack can test for 100 samples (patient, Quality Control, etc.).

Reason for recall

VITROS Immunodiagnostic Products Anti-HCV kit lot 9090 has been observed producing sporadic lower than expected VITROS Anti-HCV test results due to an approximate signal loss of up to 66.2% which can result in unexpected negative test results being obtained for reactive sample fluids (quality control and patient samples).

Data sourced from openFDA. This site is unofficial and independent of the FDA.