Recall Z-2662-2017— Exactech, Inc.

Class II · Terminated

Class II FDA medical device recall by Exactech, Inc., initiated 2017-05-25, terminated 2019-07-26. FDA's recall record names no specific 510(k) clearance for this event. Product: Logic Fit Tibial Tamp Head. Reason: The firm determined that the labeling (surgical technique) should be updated to clarify the proper technique and instrumentation to remove the tamp assembly from the bone, i.e., to include a caution statement about the potential for instrument breakage if the tamp handle/guide is misused by impacting in retro-grade..

Recalling firm
Exactech, Inc.
Classification
Class II
Status
Terminated
Initiated
2017-05-25
Terminated
2019-07-26
Firm location
Gainesville, FL, United States

Product description

Logic Fit Tibial Tamp Head

Reason for recall

The firm determined that the labeling (surgical technique) should be updated to clarify the proper technique and instrumentation to remove the tamp assembly from the bone, i.e., to include a caution statement about the potential for instrument breakage if the tamp handle/guide is misused by impacting in retro-grade.

Data sourced from openFDA. This site is unofficial and independent of the FDA.