Recall Z-2660-2023— Olympus Corporation of the Americas

Class II · Ongoing

Class II FDA medical device recall by Olympus Corporation of the Americas, initiated 2023-08-10. It names one FDA 510(k) clearance: K170023. Product: Veran Endobronchial: Models: INS-5925 SPiN Access Catheter¿ 180 Degree for use w/ Pentax Scopes INS-5920 SPiN Access Catheter¿ 90 Degree for use w/ Pentax Scopes INS-5915 SPiN Access Catheter¿ 180 for use w/ Olympus Scopes INS-5910 SPiN Access Catheter¿ 90 for use w/ Olympus 190 Scopes INS-5905 SPiN Access Catheter¿ 180 for use w/ Olympus Scopes INS-5900 SPiN Access Catheter¿ 90 Degree for use w/ Olympus Scopes INS-5700 SPiN EWC Biopsy Guide Kit. Reason: Non-conformances in Veran devices (ig4 Image Guided System, SPiN Thoracic Navigation System, SPiN Planning Laptop Workstation, and all associated navigation instruments) manufacturing to design requirements and insufficient evidence to demonstrate adequate electromagnetic compatibility (EMC). EMC is the ability of electrical equipment and systems to function acceptably without interference and to not interfere with other devices within its operating environment.

Recalling firm
Olympus Corporation of the Americas
Classification
Class II
Status
Ongoing
Initiated
2023-08-10
Firm location
Center Valley, PA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Veran Endobronchial: Models: INS-5925 SPiN Access Catheter¿ 180 Degree for use w/ Pentax Scopes INS-5920 SPiN Access Catheter¿ 90 Degree for use w/ Pentax Scopes INS-5915 SPiN Access Catheter¿ 180 for use w/ Olympus Scopes INS-5910 SPiN Access Catheter¿ 90 for use w/ Olympus 190 Scopes INS-5905 SPiN Access Catheter¿ 180 for use w/ Olympus Scopes INS-5900 SPiN Access Catheter¿ 90 Degree for use w/ Olympus Scopes INS-5700 SPiN EWC Biopsy Guide Kit

Reason for recall

Non-conformances in Veran devices (ig4 Image Guided System, SPiN Thoracic Navigation System, SPiN Planning Laptop Workstation, and all associated navigation instruments) manufacturing to design requirements and insufficient evidence to demonstrate adequate electromagnetic compatibility (EMC). EMC is the ability of electrical equipment and systems to function acceptably without interference and to not interfere with other devices within its operating environment

Data sourced from openFDA. This site is unofficial and independent of the FDA.