Recall Z-2650-2016— Medtronic Inc., Cardiac Rhythm and Heart Failure
Class II · Terminated
Class II FDA medical device recall by Medtronic Inc., Cardiac Rhythm and Heart Failure, initiated 2016-08-12, terminated 2017-01-20. FDA's recall record names no specific 510(k) clearance for this event. Product: Medtronic, Viva Quadripolar Implantable Cardioverter Defibrillators: Product Model VIVAXTCRT-D DTBA1D1, DTBA1D4 VIVASCRT-D DTBB1D1, DTBB1D4 Product Usage: The Medtronic Viva CRT -D dual chamber, implantable cardioverter defibrillator with cardiac resynchronization therapy (CRT -D) is a multiprogrammable cardiac device that monitors and regulates the patients heart rate by providing single or dual chamber, rate responsive bradycardia pacing; sequential biventricular pacing; ventricular tachyarrhythmia therapies; and atrial tachyarrhythmia therapies.. Reason: 78 Viva CRT -Ds and Evera ICDs that were manufactured with a specific subset of circuit components may experience rapid battery depletion due to a low resistance path developing within the circuit component. There have been seven (7) confirmed failures (9%) through July 16, 2016. Reported complications have included shortness of breath, pocket heating, low heart rate, and early device explant..
- Recalling firm
- Medtronic Inc., Cardiac Rhythm and Heart Failure
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2016-08-12
- Terminated
- 2017-01-20
- Firm location
- Mounds View, MN, United States
Product description
Medtronic, Viva Quadripolar Implantable Cardioverter Defibrillators: Product Model VIVAXTCRT-D DTBA1D1, DTBA1D4 VIVASCRT-D DTBB1D1, DTBB1D4 Product Usage: The Medtronic Viva CRT -D dual chamber, implantable cardioverter defibrillator with cardiac resynchronization therapy (CRT -D) is a multiprogrammable cardiac device that monitors and regulates the patients heart rate by providing single or dual chamber, rate responsive bradycardia pacing; sequential biventricular pacing; ventricular tachyarrhythmia therapies; and atrial tachyarrhythmia therapies.
Reason for recall
78 Viva CRT -Ds and Evera ICDs that were manufactured with a specific subset of circuit components may experience rapid battery depletion due to a low resistance path developing within the circuit component. There have been seven (7) confirmed failures (9%) through July 16, 2016. Reported complications have included shortness of breath, pocket heating, low heart rate, and early device explant.
Data sourced from openFDA. This site is unofficial and independent of the FDA.