Recall Z-2647-2016— Bovie Medical Corporation

Class II · Terminated

Class II FDA medical device recall by Bovie Medical Corporation, initiated 2016-07-18, terminated 2016-12-14. It names one FDA 510(k) clearance: K152570. Product: BVX-330NR Bovie Precise 360 Handpiece 33cm, needle Product is used for the delivery of helium gas plasma for cutting, coagulation, and ablation of soft tissue during open and laparoscopic surgical procedures.. Reason: The adhesive may be inadequately applied or the assembly may be manipulated after application of the adhesive resulting in an incomplete bond to the angled tip..

Recalling firm
Bovie Medical Corporation
Classification
Class II
Status
Terminated
Initiated
2016-07-18
Terminated
2016-12-14
Firm location
Clearwater, FL, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

BVX-330NR Bovie Precise 360 Handpiece 33cm, needle Product is used for the delivery of helium gas plasma for cutting, coagulation, and ablation of soft tissue during open and laparoscopic surgical procedures.

Reason for recall

The adhesive may be inadequately applied or the assembly may be manipulated after application of the adhesive resulting in an incomplete bond to the angled tip.