Recall Z-2644-2017— Exactech, Inc.

Class II · Terminated

Class II FDA medical device recall by Exactech, Inc., initiated 2017-05-22, terminated 2019-07-26. FDA's recall record names no specific 510(k) clearance for this event. Product: Exactech 1.5" Novation Calcar Planer Guide Tip To assist the surgeon in the implantation of Novation hip system components according to a conventional technique for total hip replacement. Specifically, these instruments are reusable, surgically invasive, and intended for transient use. ech 1.5" Novation Calcar Planer Guide Tip. Reason: The firm determined that the labeling (surgical technique) should be updated to clarify the proper technique and instrumentation to perform calcar planing to the bone, i.e., to include a caution statement about the potential for instrument breakage if bending or torsional forces are applied..

Recalling firm
Exactech, Inc.
Classification
Class II
Status
Terminated
Initiated
2017-05-22
Terminated
2019-07-26
Firm location
Gainesville, FL, United States

Product description

Exactech 1.5" Novation Calcar Planer Guide Tip To assist the surgeon in the implantation of Novation hip system components according to a conventional technique for total hip replacement. Specifically, these instruments are reusable, surgically invasive, and intended for transient use. ech 1.5" Novation Calcar Planer Guide Tip

Reason for recall

The firm determined that the labeling (surgical technique) should be updated to clarify the proper technique and instrumentation to perform calcar planing to the bone, i.e., to include a caution statement about the potential for instrument breakage if bending or torsional forces are applied.

Data sourced from openFDA. This site is unofficial and independent of the FDA.