Recall Z-2643-2014— Siemens Healthcare Diagnostics

Class II · Terminated

Class II FDA medical device recall by Siemens Healthcare Diagnostics, initiated 2014-08-06, terminated 2017-02-28. It names one FDA 510(k) clearance: K081895. Product: ADVIA Chemistry A1c_3 CAL (Calibrator) for use on the ADVIA Chemistry Systems, REF/Catalog Number/Siemens Material Number (SMN) 10491408, IVD ADVIA Chemistry A1c_3 Calibrator Information: For in vitro diagnostic use in the calibration of A1c_3 (A1c_3 and A1c_3M) and Total Hemoglobin_3 (tHb_3 and tHb_3M) methods on ADVIA Chemistry systems. The ADVIA Chemistry systems are automated, clinical chemistry analyzers that can run tests on human serum, plasma, or urine in random access and batch modes for in vitro diagnostic use.. Reason: The ADVIA Chemistry Systems A1c_3 method may exhibit a percent bias of -9% to +11% for hemoglobin (Hb)A1c samples when using A1c_3 Calibrator lots 3HD044 and 3LD068. Depending on quality control limits, this issue may not have been detected. NOTE: The percent bias range is a percentage of the HbA1c result. It is NOT absolute HbA1c units..

Recalling firm
Siemens Healthcare Diagnostics
Classification
Class II
Status
Terminated
Initiated
2014-08-06
Terminated
2017-02-28
Firm location
Tarrytown, NY, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

ADVIA Chemistry A1c_3 CAL (Calibrator) for use on the ADVIA Chemistry Systems, REF/Catalog Number/Siemens Material Number (SMN) 10491408, IVD ADVIA Chemistry A1c_3 Calibrator Information: For in vitro diagnostic use in the calibration of A1c_3 (A1c_3 and A1c_3M) and Total Hemoglobin_3 (tHb_3 and tHb_3M) methods on ADVIA Chemistry systems. The ADVIA Chemistry systems are automated, clinical chemistry analyzers that can run tests on human serum, plasma, or urine in random access and batch modes for in vitro diagnostic use.

Reason for recall

The ADVIA Chemistry Systems A1c_3 method may exhibit a percent bias of -9% to +11% for hemoglobin (Hb)A1c samples when using A1c_3 Calibrator lots 3HD044 and 3LD068. Depending on quality control limits, this issue may not have been detected. NOTE: The percent bias range is a percentage of the HbA1c result. It is NOT absolute HbA1c units.