Recall Z-2638-2026— Baxter Healthcare Corporation
Class II · Ongoing
Class II FDA medical device recall by Baxter Healthcare Corporation, initiated 2026-04-21. FDA's recall record names no specific 510(k) clearance for this event. Product: Adapter for Head Positioning, Model/Catalog Number: 1739994, when used in conjunction with the side rails of the Upper Back Section H V U component (product code 1909813) on Baxter operating tables. Reason: The Adapter for Head Positioning accessory may become loose during patient positioning and unintentionally move when used in conjunction with the side rails of the Upper Back Section H V U component on Baxter operating tables..
- Recalling firm
- Baxter Healthcare Corporation
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2026-04-21
- Firm location
- Deerfield, IL, United States
Product description
Adapter for Head Positioning, Model/Catalog Number: 1739994, when used in conjunction with the side rails of the Upper Back Section H V U component (product code 1909813) on Baxter operating tables
Reason for recall
The Adapter for Head Positioning accessory may become loose during patient positioning and unintentionally move when used in conjunction with the side rails of the Upper Back Section H V U component on Baxter operating tables.