Recall Z-2630-2025— PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.

Class II · Ongoing

Class II FDA medical device recall by PHILIPS MEDICAL SYSTEMS NEDERLAND B.V., initiated 2025-09-03. It names 6 FDA 510(k) clearances: K161563, K141979, K133292 and 3 more. Product: Allura Xper FD20/20; Product Code: 722038; Associated 510(k) numbers: K102005, K130842, K130638, K133292, K141979, K161563, K162859, K033737, K041949;. Reason: Systems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become unavailable. Manual movements of the stand (longitudinal, transversal and rotational for ceiling mounted systems and rotational for floor mounted systems) and table longitudinal movements remain available. Imaging (X-ray) functionality remains available..

Recalling firm
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Classification
Class II
Status
Ongoing
Initiated
2025-09-03
Firm location
Best, N/A, Netherlands

510(k) clearances named in this recall

6 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Allura Xper FD20/20; Product Code: 722038; Associated 510(k) numbers: K102005, K130842, K130638, K133292, K141979, K161563, K162859, K033737, K041949;

Reason for recall

Systems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become unavailable. Manual movements of the stand (longitudinal, transversal and rotational for ceiling mounted systems and rotational for floor mounted systems) and table longitudinal movements remain available. Imaging (X-ray) functionality remains available.

Data sourced from openFDA. This site is unofficial and independent of the FDA.