Recall Z-2616-2016— Cook Inc.
Class II · Terminated
Class II FDA medical device recall by Cook Inc., initiated 2016-04-15, terminated 2017-08-04. FDA's recall record names no specific 510(k) clearance for this event. Product: Haskal Transjugular Intrahepatic Portal Access Set Introducer, catheter Intended for transjugular liver access in diagnostic and interventional procedures.. Reason: Increase in reports of polymer degradation of the catheter tip, resulting in tip fracture and/or separation. degradation could include loss of device function, separation of a device segment leading to medical intervention, or complications resulting from a separated segment. Such complications include device fragments in the vascular system, genitourinary system, or other soft tissues. Fragments within the vascular system could result in embolization to the heart or lungs, or occluding blood flow to end organs..
- Recalling firm
- Cook Inc.
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2016-04-15
- Terminated
- 2017-08-04
- Firm location
- Bloomington, IN, United States
Product description
Haskal Transjugular Intrahepatic Portal Access Set Introducer, catheter Intended for transjugular liver access in diagnostic and interventional procedures.
Reason for recall
Increase in reports of polymer degradation of the catheter tip, resulting in tip fracture and/or separation. degradation could include loss of device function, separation of a device segment leading to medical intervention, or complications resulting from a separated segment. Such complications include device fragments in the vascular system, genitourinary system, or other soft tissues. Fragments within the vascular system could result in embolization to the heart or lungs, or occluding blood flow to end organs.
Data sourced from openFDA. This site is unofficial and independent of the FDA.