Recall Z-2615-2026— Medline Industries, LP
Class I · Ongoing
Class I FDA medical device recall by Medline Industries, LP, initiated 2026-05-15. It names one FDA 510(k) clearance: K241156. Product: Reprocessed Electrophysiology Catheters and Ultrasound Catheters Item no. 401206RH, St. Jude Medical Response" Diagnostic Electrophysiology Catheter 4 Elec. 10mm; Item no. 401305RH, St. Jude Medical Response" Diagnostic Electrophysiology Catheter 10 Elec. 2-5-2mm.. Reason: Expansion of RES 98277. Specific lots of reprocessed catheters may contain small particles of residual material. If devices with residual particulates on patient-contacting surfaces are used, there is a potential risk of an inflammatory response or systemic infection. If particulate were to dislodge during use and contact circulating blood or the heart, it may produce a granulomatous reaction or thrombus formation, which may lead to cerebral or pulmonary embolism, or deep vein thrombosis..
- Recalling firm
- Medline Industries, LP
- Classification
- Class I
- Status
- Ongoing
- Initiated
- 2026-05-15
- Firm location
- Northfield, IL, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Reprocessed Electrophysiology Catheters and Ultrasound Catheters Item no. 401206RH, St. Jude Medical Response" Diagnostic Electrophysiology Catheter 4 Elec. 10mm; Item no. 401305RH, St. Jude Medical Response" Diagnostic Electrophysiology Catheter 10 Elec. 2-5-2mm.
Reason for recall
Expansion of RES 98277. Specific lots of reprocessed catheters may contain small particles of residual material. If devices with residual particulates on patient-contacting surfaces are used, there is a potential risk of an inflammatory response or systemic infection. If particulate were to dislodge during use and contact circulating blood or the heart, it may produce a granulomatous reaction or thrombus formation, which may lead to cerebral or pulmonary embolism, or deep vein thrombosis.
Data sourced from openFDA. This site is unofficial and independent of the FDA.