Recall Z-2614-2016— Cook Inc.
Class II · Terminated
Class II FDA medical device recall by Cook Inc., initiated 2016-04-15, terminated 2017-08-04. It names one FDA 510(k) clearance: K122937. Product: Slip-Cath Beacon Tip Catheter Catheter, Intravascular, Diagnostic The catheters are intended for use in angiographic procedures by physicians trained and experienced in angiographic techniques.. Reason: Increase in reports of polymer degradation of the catheter tip, resulting in tip fracture and/or separation. degradation could include loss of device function, separation of a device segment leading to medical intervention, or complications resulting from a separated segment. Such complications include device fragments in the vascular system, genitourinary system, or other soft tissues. Fragments within the vascular system could result in embolization to the heart or lungs, or occluding blood flow to end organs..
- Recalling firm
- Cook Inc.
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2016-04-15
- Terminated
- 2017-08-04
- Firm location
- Bloomington, IN, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Slip-Cath Beacon Tip Catheter Catheter, Intravascular, Diagnostic The catheters are intended for use in angiographic procedures by physicians trained and experienced in angiographic techniques.
Reason for recall
Increase in reports of polymer degradation of the catheter tip, resulting in tip fracture and/or separation. degradation could include loss of device function, separation of a device segment leading to medical intervention, or complications resulting from a separated segment. Such complications include device fragments in the vascular system, genitourinary system, or other soft tissues. Fragments within the vascular system could result in embolization to the heart or lungs, or occluding blood flow to end organs.
Data sourced from openFDA. This site is unofficial and independent of the FDA.