Recall Z-2612-2016— Cook Inc.

Class II · Terminated

Class II FDA medical device recall by Cook Inc., initiated 2016-04-15, terminated 2017-08-04. FDA's recall record names no specific 510(k) clearance for this event. Product: Beacon Tip Centimeter Sizing Catheter Beacon Tip White Vessel Sizing Catheter Beacon Tip Vessel Sizing Catheter Catheter, Intravascular, Diagnostic The catheters are intended for use in angiographic procedures by physicians trained and experienced in angiographic techniques. Standard techniques for placement of vascular access sheaths, angiographic catheters and wire guides should be employed.. Reason: Increase in reports of polymer degradation of the catheter tip, resulting in tip fracture and/or separation. degradation could include loss of device function, separation of a device segment leading to medical intervention, or complications resulting from a separated segment. Such complications include device fragments in the vascular system, genitourinary system, or other soft tissues. Fragments within the vascular system could result in embolization to the heart or lungs, or occluding blood flow to end organs..

Recalling firm
Cook Inc.
Classification
Class II
Status
Terminated
Initiated
2016-04-15
Terminated
2017-08-04
Firm location
Bloomington, IN, United States

Product description

Beacon Tip Centimeter Sizing Catheter Beacon Tip White Vessel Sizing Catheter Beacon Tip Vessel Sizing Catheter Catheter, Intravascular, Diagnostic The catheters are intended for use in angiographic procedures by physicians trained and experienced in angiographic techniques. Standard techniques for placement of vascular access sheaths, angiographic catheters and wire guides should be employed.

Reason for recall

Increase in reports of polymer degradation of the catheter tip, resulting in tip fracture and/or separation. degradation could include loss of device function, separation of a device segment leading to medical intervention, or complications resulting from a separated segment. Such complications include device fragments in the vascular system, genitourinary system, or other soft tissues. Fragments within the vascular system could result in embolization to the heart or lungs, or occluding blood flow to end organs.

Data sourced from openFDA. This site is unofficial and independent of the FDA.