Recall Z-2611-2015— Customed, Inc

Class II · Terminated

Class II FDA medical device recall by Customed, Inc, initiated 2015-07-10, terminated 2017-04-12. FDA's recall record names no specific 510(k) clearance for this event. Product: MAYO STAND COVER W/CSR WRAP 30/CS Used by medical staff to perform surgical procedures and patient care procedures such as wound healing, suture removals, irrigate wounds, etc. .. Reason: Recall initiated due to inadequate validation/qualification of sterilization processes, and package integrity for products released for distribution as sterile..

Recalling firm
Customed, Inc
Classification
Class II
Status
Terminated
Initiated
2015-07-10
Terminated
2017-04-12
Firm location
Fajardo, PR, United States

Product description

MAYO STAND COVER W/CSR WRAP 30/CS Used by medical staff to perform surgical procedures and patient care procedures such as wound healing, suture removals, irrigate wounds, etc. .

Reason for recall

Recall initiated due to inadequate validation/qualification of sterilization processes, and package integrity for products released for distribution as sterile.

Data sourced from openFDA. This site is unofficial and independent of the FDA.