Recall Z-2610-2017— Siemens Healthcare Diagnostics, Inc.

Class II · Terminated

Class II FDA medical device recall by Siemens Healthcare Diagnostics, Inc., initiated 2017-03-21, terminated 2018-08-03. FDA's recall record names no specific 510(k) clearance for this event. Product: Dimension Vista Prolactin (PRL) Flex reagent cartridge. Reason: The concentrations for Biotin listed in the Non-Interfering Substances section of the current Dimension and Dimension Vistat Instructions for Use (IFU) for Dimension TNI, Dimension Vista DIGXN, E2, FERR, PRL, and TSH have incorrect units, and incorrectly state the level at which biotin does not interfere..

Recalling firm
Siemens Healthcare Diagnostics, Inc.
Classification
Class II
Status
Terminated
Initiated
2017-03-21
Terminated
2018-08-03
Firm location
Newark, DE, United States

Product description

Dimension Vista Prolactin (PRL) Flex reagent cartridge

Reason for recall

The concentrations for Biotin listed in the Non-Interfering Substances section of the current Dimension and Dimension Vistat Instructions for Use (IFU) for Dimension TNI, Dimension Vista DIGXN, E2, FERR, PRL, and TSH have incorrect units, and incorrectly state the level at which biotin does not interfere.

Data sourced from openFDA. This site is unofficial and independent of the FDA.